NejSi-3M implantable neurostimulator. Description.

NejSi-3M.ТС neurostimulator for test stimulationNeurostimulator for test stimulation
НейСі-3М.ТС

Purpose:

  • The stimulator is intended for test and intraoperative electrical stimulation using multi-contact electrodes. The electrodes connect to the stimulator through an adapter cable with a plug.
  • The stimulator delivers stimulation to each patient according to an individual programme held in non-volatile memory.
  • The stimulator allows stimulation modes and parameters to be changed and stored in non-volatile memory individually for each patient.
  • Connecting the stimulator to a personal computer over a wireless Bluetooth link allows patient stimulation settings to be edited on the computer, saved to external storage, and transferred from external storage back into the stimulator.

Main technical data and specifications:

  • The stimulator generates pulses of positive or negative polarity, and can also generate a pulse train whose polarity alternates from pulse to pulse.
  • Number of contacts – up to 4
  • The stimulator delivers stimulation either as voltage pulses or as current pulses, in "Voltage" and "Current" modes respectively. Pulses are applied to any chosen pair of contacts. They are delivered in bursts with programmable burst duration and inter-burst interval
  • Stimulation modes and parameters can be changed and stored in the stimulator's non-volatile memory separately for each patient. Number of patients: 100
  • With the battery fully charged, the stimulator runs continuously for at least 6 hours
  • The stimulator meets the electrical safety requirements of ДСТУ ГОСТ 12.2.025 – 76